Open Access Open Access  Restricted Access Subscription Access

AMS 14C Sample Preparation at the KCCAMS/UCI Facility: Status Report and Performance of Small Samples

G M Santos, R B Moore, J R Southon, S Griffin, E Hinger, D Zhang

Abstract


We present an overview of accelerator mass spectrometry (AMS) radiocarbon sample preparation and measurements, describing the technical upgrades that now allow us to routinely obtain 0.2-0.3% precision for 1-mg carbon samples. A precision of ~1% on samples with 100 g of carbon can also be achieved. We have also developed graphitization techniques and AMS procedures for ultra-small samples (down to 0.002 mg of carbon). Detailed time series are presented for large and small aliquots of standards such as NIST OX-I and OX-II; FIRI-C and -D; IAEA-C6, -C7 and -C8; and 14C-free samples.

Full Text:

PDF