SIMULTANEOUSLY MEASURING 14C AND RADON IN BENZENE DATING SAMPLES
Abstract
After benzene synthesis, radiocarbon dating samples are usually stored for 34 weeks before counting to allow
an eventual radon contamination to decay to a negligible level. This paper presents a technique that can minimize, and often
eliminate, this delay by using a simple single-phototube liquid scintillation counting system, specifically designed for 14C
dating. Radon contamination is assessed by pulses of 214Po (a 222Rn decay product, half-life 0.16 s), identified through
pulse-time analysis. For each 214Po pulse, 0.49 beta particle pulses of 214Pb and 214Bi fall in the 14C counting window, and the
214Po pulses are used to correct the 14C count rate. A 14C sample (count rate 11.6 cpm) was measured continuously for 16 days.
It was then doped with radon, which increased the first 24-hr count rate in the 14C channel by 3.8 cpm, and the sample was
measured for 27 more days. Radon did not measurably affect the 14C-corrected count rate. Counting a sample for 2 min
reveals whether it needs storing. If the radon concentration is low, the sample can be measured immediately without degrading
accuracy.
an eventual radon contamination to decay to a negligible level. This paper presents a technique that can minimize, and often
eliminate, this delay by using a simple single-phototube liquid scintillation counting system, specifically designed for 14C
dating. Radon contamination is assessed by pulses of 214Po (a 222Rn decay product, half-life 0.16 s), identified through
pulse-time analysis. For each 214Po pulse, 0.49 beta particle pulses of 214Pb and 214Bi fall in the 14C counting window, and the
214Po pulses are used to correct the 14C count rate. A 14C sample (count rate 11.6 cpm) was measured continuously for 16 days.
It was then doped with radon, which increased the first 24-hr count rate in the 14C channel by 3.8 cpm, and the sample was
measured for 27 more days. Radon did not measurably affect the 14C-corrected count rate. Counting a sample for 2 min
reveals whether it needs storing. If the radon concentration is low, the sample can be measured immediately without degrading
accuracy.